Reason stated in the source

Internal components of the Protein S ELISA Antigen Kit (Lot: 3-16-5292) were mislabeled. The Protein S Conjugate Solution is labeled as Substrate, and the Substrate is labeled as Protein S conjugate.

Source facts

Action
On 4/4/17, an email notification was initially sent to the firm's OUS distributor that received 3 kits. On 4/10/17, the customer replied with a photograph to confirm their kits were affected. A Withdrawal notice was approved internally, and formally issued by the firm to a single distributor via e-mail on 4/17/17. An authorized confirmation of destruction was sent from the distributor to the firm via e-mail on 4/18/17. The firm's public contact may be reached at (409) 842-3714 ext. 1177.
Initiated
2017-04-04
Root cause
Labeling mix-ups
Terminated
2020-05-14
Distribution
International distribution to country of: Jordan
Recall number
Z-0106-2019
Recall status
Terminated
CDRH record ID
167256
FDA record created
2018-10-12
Recall event number
81024

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗