Skip to content

Public FDA record

Z-0106-2021: Pediatric Nephrostomy Stent Set, Reference Part Numbers 080208-S6 and 080208-S7, Order Numbers G15034 and

Recall record Z-0106-2021 for Cook Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0106-2021
Company in FDA record
Cook Inc.
FDA record creation date
Oct 14, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The indwell time for the Percutaneous Neonatal Pigtail Nephrostomy Set and Pediatric Nephrostomy Stent Set is decreasing from 4 months to 4 weeks. The current IFU is in process of being updated and will be provided with orders placed following implementation.

Source facts

Action
URGENT: MEDICAL DEVICE CORRECTION notification letters dated 8/31/20 were sent to customers. Actions to be Taken by the Customer 1. If affected products have already been placed within a patient, the indwell time should not exceed 4 weeks. If the current indwell time has exceeded 4 weeks, consider removing and/or replacing the product. 2. Understand that if an affected product is placed within a patient longer than 4 weeks, there is an increased potential for breakage, kinking, encrustation, migration, and/or infection to occur. 3. Please maintain a copy of this notice with the current IFU or product(s) in your inventory. 4. If you have affected product in inventory, you may continue to use these products. Cook Medical is not requesting product be returned. 5. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product(s) on hand, you must still complete the Acknowledgement and Receipt Form and return via fax (812.339.7316) or email (FieldActionsNA@CookMedical.com). 6. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 7. Immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: CustomerRelationsNA@CookMedical.com. Action being Taken by Cook Medical Cook Medical is initiating a voluntary correction to inform customers that the indicated indwell time is changing for the Percutaneous Neonatal Pigtail Nephrostomy Set and Pediatric Nephrostomy Stent Set. The current IFU is in process of being updated and will be provided with orders placed following implementation. We recognize that this situation is a potential disruption to your normal operations, and we sincerely apologize. Thank you for your
Initiated
2020-08-31
Root cause
No Marketing Application
Terminated
2021-07-01
Distribution
Worldwide distribution - US Nationwide.
Recall number
Z-0106-2021
Recall status
Terminated
CDRH record ID
183657
FDA record created
2020-10-14
Recall event number
86411

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗