Open original FDA record ↗Recall record
- FDA identifier
- Z-0106-2023
- Company in FDA record
- MEDLINE INDUSTRIES, LP - Northfield
- FDA classification date
- Oct 20, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
One lot of chest tube insertion trays, containing a silk suture component that is restricted to 1 cycle of sterilization, was processed though 2 cycles.
Source facts
- Action
- Medline Industries issued a Medical Device Recall note on 09/08/2022 by mail and email. The notice explained the problem and the potential hazard and requested the following: "REQUIRED ACTION: 1.Immediately check your stock for the affected item number and the affected lot numbers listed on the recall portal. Destroy affected product. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. The login for completing the response form is: Website link: https://recalls.medline.com Recall Reference #: R-22-131-FG Recall Code: RXXXXXXX 3.Upon completion of the form, please destroy affected product. Your account will receive credit once the response form is submitted. 4.If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Have the customers document and destroy any affected product. You should include your customers quantities on your response form. If you have any questions, please contact the Recall Department at 866-359-1704."
- Initiated
- 2022-09-07
- Root cause
- Under Investigation by firm
- Terminated
- 2026-03-04
- Distribution
- US Nationwide distribution in the states of CA and TX.
- Recall number
- Z-0106-2023
- Recall status
- Terminated
- CDRH record ID
- 196170
- FDA classified
- 2022-10-20
- Recall event number
- 90901