Open original FDA record ↗Recall record
- FDA identifier
- Z-0107-2012
- Company in FDA record
- GE Healthcare, LLC
- FDA classification date
- Oct 26, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
GE Healthcare has recently become aware that an exposed high voltage gradient wire internal to the MSK 1.5T Extreme system equipment may cause electrical shock, which may impact service personnel safety. This issue does not impact patient or operator safety.
Source facts
- Action
- GE Healthcare LLC sent a "Urgent Medical Device Correction letter" dated August 29, 2011, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The customer was instructed to please ensure that all potential service personnel at your facility are made aware of this safety notification and the recommended actions. Contact your local service representative if you have any questions concerning this notification. For further questions please call ( 262 ) 513-4122.
- Initiated
- 2011-08-29
- Root cause
- Nonconforming Material/Component
- Terminated
- 2012-05-29
- Distribution
- Worldwide Distribution -- USA (nationwide) including the states of AL, CA, FL, ID, NY, NH, NJ, NV, OH, PA, TX, VA and WI., and countries of UNITED ARAB EMIRATES, UKRAINE, UNITED KINGDOM, SWITZERLAND
- Recall number
- Z-0107-2012
- Recall status
- Terminated
- CDRH record ID
- 103010
- FDA classified
- 2011-10-26
- Recall event number
- 59522