Specific Pumps may exhibit intermittent performance in the AILD (Air in Line Detector) function. If an AILD does not sufficiently discern fluid from air in line, an air in line event may not be recognized by the pump and may not alarm to notify the clinician.
Source facts
Action
The firm notified its consignees by email beginning 03/24/2020. The notice explained the issue, risk to health, and requested a response. Distributors/pharmacy suppliers were directed to notify their customers. The firm is seeking return of the units for correction.
Initiated
2020-03-24
Root cause
Component change control
Terminated
2023-12-22
Distribution
Distribution to European countries: BE, BH, CH, DE, DK, ES, FI, FR, GB, IT, HL, NO, PT, SE
Recall number
Z-0107-2022
Recall status
Terminated
CDRH record ID
189372
FDA classified
2021-10-19
Recall event number
88701