Z-0107-2024: DxI 9000 Access Immunoassay Analyzer, REF C11137, Sofware Versions 1.15 and below; invitro diagnostic

Recall record Z-0107-2024 for Beckman Coulter, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0107-2024
Company in FDA record
Beckman Coulter, Inc.
FDA classification date
Oct 12, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm has investigated and confirmed that when DxI 9000 is connected to Laboratory Information System (LIS, host), the sample barcode presented may be ignored and replaced with a sample ID that was previously in the same rack and position. This will cause a delay of result to the sample currently presented to the system (Sample B) and may result in an erroneous test result for the prior sample (Sample A), which gets associated with the results of sample B.

Source facts

Action
Beckman Coulter notified its consignees on 09/12/2023 via letter. The notice explained the issue, risk, and consignees were advised to perform a retrospective review of all results that were generated on a DxI 9000 analyzer that is connected to a host system (laboratory information system or middleware). Current customers should perform one of the following: Option 1: Disconnect the DxI 9000 Access Immunoassay Analyzer from the host system. You may continue ordering tests through the analyzer console. Option 2: Prevent the host system from downloading test orders or sending new information updates to the analyzer. Beckman Coulter will release system software version 1.15.2, which prevents reporting an erroneous test result. Test results will still be delayed when the DxI 9000 Access Immunoassay Analyzer may incorrectly identify samples.
Initiated
2023-09-12
Root cause
Software design
Distribution
Worldwide - US Nationwide distribution in the state of Florida, and the countries of Croatia, Czechia, France, Germany, Ireland, Italy, Netherlands, New Zealand, South Africa, Spain, Switzerland, United Kingdom of Great Britain and Northern Ireland.
Recall number
Z-0107-2024
Recall status
Open, Classified
CDRH record ID
203296
FDA classified
2023-10-12
Recall event number
93023

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗