Skip to content

Public FDA record

Z-0108-2012: Nebulizer (adult elastic strap style aerosol mask 7' tube - 50/cs), Latex Free Salter Labs, Arvin, CA 932

Recall record Z-0108-2012 for Salter Laboratories, Division of Regulatory Affairs. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0108-2012
Company in FDA record
Salter Laboratories, Division of Regulatory Affairs
FDA classification date
Oct 26, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

A small number of oxygen mask components were found to have 100% obstructed airways rendering them incapable of delivering oxygen to patients undergoing respiration therapy.

Source facts

Action
The firm, Salter Labs, called all their customers on March 25, 2011 and asked that they hold all affected product. In addition, the firm sent an "URGENT PRODUCT RECALL" letter dated April 4, 2011 and a follow-up letter dated April 7, 2011 via e-mail and certified mail to all customers. The letter described the product, problem and actions to be taken. The customers were instructed to examine their inventory, return any affected lots listed, contact Salter Customer Service at 1-800-235-4203 to arrange for return of affected product and complete and return the Customer Reply Form via 661-854-6816 or toll free 1-800-628-4690. Should you have any questions, please contact Salter Labs at 1-800-235-4203.
Initiated
2011-03-25
Root cause
Equipment maintenance
Terminated
2011-10-26
Distribution
Worldwide distribution: USA (nationwide) and countries of: Belgium, Canada, Colombia and Iceland.
Recall number
Z-0108-2012
Recall status
Terminated
CDRH record ID
99387
FDA classified
2011-10-26
Recall event number
58454

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗