Open original FDA record ↗Recall record
- FDA identifier
- Z-0108-2015
- Company in FDA record
- St Jude Medical
- FDA classification date
- Oct 22, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
St. Jude Medical is conducting a voluntary recall of SMJ Batch # 4641580 of the Fast-Cath Trio" Hemostasis Introducer, Catalog number 406308. The inner diameter of the 10F dilator is too small, such that the guidewire cannot advance through the device which may result in a procedural delay while another dilator of the desired size is obtained.
Source facts
- Action
- St. Jude Medical sent an Urgent Medical Device Recall Notice dated October 14, 2014, to all affected consignees. The letter described the problem and the product affected by the recall. Advised consignees to discontinue us of the device. A St. Jude Medical sales representative were to contact the consignee in the near future to facilitate removal and reconciliation of the recalled device. Customers with questions were instructed to contact their St. Jude Medical sales Representative. For questions regarding this recall call 651-756-2295.
- Initiated
- 2014-10-14
- Root cause
- Process control
- Terminated
- 2015-03-25
- Distribution
- Nationwide Distribution including FL, IN, ME, MA, NY, NY, OH, RI,TX, and WV.
- Recall number
- Z-0108-2015
- Recall status
- Terminated
- CDRH record ID
- 130765
- FDA classified
- 2014-10-22
- Recall event number
- 69499