Z-0108-2019: Carl Zeiss Meditec AG, Goeschwitzer Strasse 51-52 07745 Jena, Germany, IOLMaster 700, REF 1932-169 SN 118

Recall record Z-0108-2019 for Carl Zeiss Meditec AG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0108-2019
Company in FDA record
Carl Zeiss Meditec AG
FDA record creation date
Oct 15, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Calibration setting was incorrectly entered during manufacturing causing the optical coherence tomography (OCT) to obtain the incorrect two-dimensional images of ocular structures of the anterior and posterior segments of the eye.

Source facts

Action
On 08/06/2018 the firm sent a Field Safety Corrective Action letter to customers via FedEx Priority Overnight service informing them that based on routine post market surveillance activities, we have information that this specific IOLMaster 700 device is incorrectly calibrated. The calibration value for axial length measurement was not entered correctly in the device settings during the manufacturing process. The transmission discrepancy leads to an error during the axial length measurement of approx. 0.25mm. A deviation in the axial length of this magnitude results in an error of Single unit. approx. 0.8 Diopter with regard to the intraocular lens. This in turn results in a refractive error of about 0.5 Diopter. The local firm's service organization will remove this device from the field. The device manufacturer will recalibrate the device to ensure it meets specifications. Hazard involved: The likelihood of a faulty entered calibration value for axial length measurement in the production process of IOLMaster 700 is considered rare. The probability of injury, i.e. reduced visual acuity due to the implantation of a wrong IOL power, is very likely for the affected device, because the axial length of the eye cannot be measured correctly. The device failure may only be recognized through post-surgical examination. Affected product: Quantity: 1 Product Name: IOLMaster 700 Product Number: 000000-1932-169 Device Serial Number: 1185393 Manufacturing Date: 2017-09-18 Actions to be taken: 1. Discontinue use of the ZEISS IOLMaster 700 device (Serial Number 1185393). Physicians are asked to review refractive results during usage of this device and to inform the firm about any hyperopic refractive surprises noticed. Any measurement and calculation data from this device should not be taken into consideration for further treatments. 2. Complete and return the attached Medical Device Recall Return Response Form to USFCA.meditec.us@zeiss.com no later than Jun
Initiated
2018-06-25
Root cause
Employee error
Terminated
2021-01-15
Distribution
US: CA (Mountain View and San Jose) A single system. No distributed OUS.
Recall number
Z-0108-2019
Recall status
Terminated
CDRH record ID
167198
FDA record created
2018-10-15
Recall event number
81000

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗