Z-0108-2021: Artis zee or Artis Q systems (listed below) - Product Usage: for single and biplane diagnostic imaging an

Recall record Z-0108-2021 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0108-2021
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA record creation date
Oct 15, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential exists for the collision supervision not to work properly following loss of the individual room configuration data settings causing danger of collision with fix mounted room equipment, walls or floor may occur. This might cause components of the system to fall or tilt which could result in crushing of patients, operators or staff, collision of system parts with patients, as well as result in delay or interruption of the clinical procedure

Source facts

Action
Siemens iMedical Solutions USA Inc. issued A Customer Safety Advisory Notice dated 8/03/20 affected customers via AX047/20/S. Letter states reason for recall, health risk and action to take:Siemens strongly recommends paying close attention during system movements to avoid any collision and to establish appropriate emergency procedures until the corrective action has been performed. In case of collision, it is recommended to check the system for any damages before continuing with treatment. If any device part continues to move although that movement was not released, a potential malfunction may be present. In this case, the system must be shut down and Siemens Healthineers service organization must be informed. Our service organization will contact you to schedule an appointment to perform inspection of the affected system by Siemens service engineer, and, where necessary, recover the individual room configuration data setings This correction will be implemented via Update Instruction AX046/20/S. Following the inspection by a service engineer the individual room configuration data settings will be available again. Please feel free to contact our service organization at 1-800-888-7436 for an earlier appointment. Following the correction, the cause will be eliminated, and any recurrence of this potential fault is prevented. Acknowledge Receipt of this Safety Advisory Notice Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein.
Initiated
2020-08-03
Root cause
Software Manufacturing/Software Deployment
Terminated
2021-02-03
Distribution
US Natonwide distribution.
Recall number
Z-0108-2021
Recall status
Terminated
CDRH record ID
183198
FDA record created
2020-10-15
Recall event number
86250

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗