Open original FDA record ↗Recall record
- FDA identifier
- Z-0108-2027
- Company in FDA record
- Canon Medical System, USA, INC.
- FDA classification date
- Oct 7, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Loosening of the bolts securing the longitudinal movement mechanism may cause the tabletop to shift, resulting in [positional/imaging/patient] misalignment and a longitudinal motion error
Source facts
- Action
- On August 26, 2026, Canon Medical Systems USA, Inc. issued a Urgent: Medical Device Correction notification to affected consignees via mail. Canon ask consignees to take the following actions: 1. Should this problem occur before inspection of the tightening of the bolt(s) is implemented on your system, please stop using the device and call your service representative. 2. Additionally, it is strongly requested that you share the contents of this letter with all users as well as clinical engineering or biomedical group at your facility. 3. A representative of Canon Medical Systems USA will contact you to schedule a time to inspect the tightening and properly fixing of the bolt(s) to correct this issue with your system(s). 4. If you have any questions regarding this matter, please contact your service representative contact our InTouch team at (800) 521-1968 or by email at intouchdispatch@us.medical.canon. 5. Canon respectfully requests that the attached form be completed and returned via email to fieldcorrections@us.medical.canon.
- Initiated
- 2026-08-26
- Root cause
- Process control
- Distribution
- US Nationwide distribution in the state of ID.
- Recall number
- Z-0108-2027
- Recall status
- Open, Classified
- CDRH record ID
- 222526
- FDA classified
- 2026-10-07
- Recall event number
- 99817