Reason stated in the source
Test Cartridges may leak and have the potential to generate error messages and biased inaccurate results.
Source facts
- Action
- BIO-RAD Laboratories Deeside sent an "URGENT: MEDICAL DEVICE CORRECTION" letter dated September 7, 2010, via registered mail to all subsidiaries/customers. The letter described the product, problem and actions to be taken by the customer. The customers were instructed to stop using and discard the test cartridges from the listed lots; complete and return the attached CUSTOMER MEDICAL DEVICE CORRECTION RESPONSE FORM via fax +44 1244 833401; record the number of affected kits of product, confirm that they have discarded the cartridges and record the method of disposal. If you have any questions, contact +44 1244 288888.
- Initiated
- 2010-09-07
- Root cause
- Component design/selection
- Terminated
- 2011-05-17
- Distribution
- Worldwide distribution: USA and countries including: Australia, New Zealand, Taiwan, France, the UK, Singapore, Germany, Italy, Thailand, China, Korea, Sweden, Sri Lanka, Maldives, Canada and the Philippines.
- Recall number
- Z-0109-2011
- Recall status
- Terminated
- CDRH record ID
- 94466
- FDA classified
- 2010-10-21
- Recall event number
- 56762