Open original FDA record ↗Recall record
- FDA identifier
- Z-0109-2019
- Company in FDA record
- Besmed Health Business Corporation
- FDA record creation date
- Oct 15, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Ambu Aura40 Standard, Reusable Laryngeal Mask packaged in a pouch marked AuraStraight Silicone, Single Use Laryngeal Mask Sterile
Source facts
- Action
- Ambu issued via USPS Certified Mail/Email to customers on Aug 09, 2018 advising users of the problem. If you have affected lot in stock and you are using it as Reusable Laryngeal Mask (following accompanying instruction for use), you can continue to use it and no need to return it. Accounts requested to see if you have affected lots in stock. If you do have it, check Aura40 Standard devices for incorrect labelled pouch, and place wrongly labelled products in quarantine. For collection and replacement by Ambu..Furthermore, please instruct all customers of the received LOT(s), to check their devices for the wrong labelling as well Contact 800 262 8462 x 1121 or email mlo@ambu.com to arrange Return Authorization and replacement.
- Initiated
- 2018-08-09
- Root cause
- Error in labeling
- Terminated
- 2019-05-10
- Distribution
- US Nationwide, and country of Canada.
- Recall number
- Z-0109-2019
- Recall status
- Terminated
- CDRH record ID
- 167221
- FDA record created
- 2018-10-15
- Recall event number
- 81009