Open original FDA record ↗Recall record
- FDA identifier
- Z-0109-2021
- Company in FDA record
- Skytron, Div. The KMW Group, Inc
- FDA record creation date
- Oct 15, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Intermittent failure of a signal being sent to a monitor from the touch panel display causing the displaying monitor to go Blank and the touch panel to display a Blank image when reviewing the target display. If this issue were to happen during a case, cause increase risk to patient due to hospital losing the visual image.
Source facts
- Action
- Skytron Authorized Representatives contacted the affected facilities beginning on 3/4/20 via phone or in person .Skytron Authorized Representatives visited sites to correct the issue wih the following message: Skytron, LLC is voluntarily recalling a limited number of SDS integration units. The SDS router may experience an intermittent issue related to display output functionality. This failure poses a potential safety risk to patients. This firmware update is being conducted to correct the failure.
- Initiated
- 2020-03-10
- Root cause
- Software Manufacturing/Software Deployment
- Terminated
- 2021-06-01
- Distribution
- US Nationwide distribution including in the states of CA, FL, IA, MI, MN, NY, PA, TX, WA.
- Recall number
- Z-0109-2021
- Recall status
- Terminated
- CDRH record ID
- 183711
- FDA record created
- 2020-10-15
- Recall event number
- 86439