Skip to content

Public FDA record

Z-0109-2021: SkyVision SDS System - Video Integration Model: GC-SDS-BSCB-200 - Product Usage: is a high quality, self-

Recall record Z-0109-2021 for Skytron, Div. The KMW Group, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0109-2021
Company in FDA record
Skytron, Div. The KMW Group, Inc
FDA record creation date
Oct 15, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Intermittent failure of a signal being sent to a monitor from the touch panel display causing the displaying monitor to go Blank and the touch panel to display a Blank image when reviewing the target display. If this issue were to happen during a case, cause increase risk to patient due to hospital losing the visual image.

Source facts

Action
Skytron Authorized Representatives contacted the affected facilities beginning on 3/4/20 via phone or in person .Skytron Authorized Representatives visited sites to correct the issue wih the following message: Skytron, LLC is voluntarily recalling a limited number of SDS integration units. The SDS router may experience an intermittent issue related to display output functionality. This failure poses a potential safety risk to patients. This firmware update is being conducted to correct the failure.
Initiated
2020-03-10
Root cause
Software Manufacturing/Software Deployment
Terminated
2021-06-01
Distribution
US Nationwide distribution including in the states of CA, FL, IA, MI, MN, NY, PA, TX, WA.
Recall number
Z-0109-2021
Recall status
Terminated
CDRH record ID
183711
FDA record created
2020-10-15
Recall event number
86439

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗