Skip to content

Public FDA record

Z-0110-06: The Lysus Infusion System - Catheter Interface Cable. The Lysus Infusion System employs high frequency (

Recall record Z-0110-06 for Ekos Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0110-06
Company in FDA record
Ekos Corp
FDA classification date
Nov 1, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The 100V board in the catheter interface cable pod failed during use. The PT-3 control unit would not recognize the catheter interface cable.

Source facts

Action
On 9/21/04 the firm began visiting customers and replacing the catheter interface cables.
Initiated
2004-09-21
Root cause
Other
Terminated
2005-11-08
Distribution
Units were shipped to hospitals, one each in WA, AZ, and OH.
Recall number
Z-0110-06
Recall status
Terminated
CDRH record ID
41800
FDA classified
2005-11-01
Recall event number
33394

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗