Skip to content

Public FDA record

Z-0110-2022: United Orthopedics USTAR II Knee System USTAR II Knee System Segment Part, RHS, 25 mm Length, Catalog #

Recall record Z-0110-2022 for Uoc Usa Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0110-2022
Company in FDA record
Uoc Usa Inc
FDA record creation date
Oct 14, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to product not meeting specification for concentricity and failing to meet requirements of dynamic fatigue testing.

Source facts

Action
On 08/27/2021, the Recalling Firm communicated via email a letter to customers/US surgeons informing them that they are initiating a product recall limited to its 25 mm segment from its knee system due to an out of specification condition that has the potential to pose a risk to patients. Customers are instructed to return unused products. For questions, contact the Director of Customer Service at 949-328-3366 or the Vice President of Marketing at 949-501-8575. On 09/16/2021, the firm sent an "URGENT: MEDICAL DEVICE RECALL" notification letter to additional customers, who have not used or received any of the affected product, just to make them aware of the recall. The Urgent Medical Device Recall notification identified the reason for the recall, the potential risks associate with the use of the affected product, and instructed these customers that to notify the recalling firm is they have any of the affected products. The Recalling Firm can be contacted via email harsh.gupta@unitedorthopedic.com or via phone at 949-501-8575. On 09/24/2021, the Recalling Firm communicated an updated "URGENT: MEDICAL DEVICE RECALL" Notification Letter informing customers that had received the affected recalled devices, of the recall action and the potential hazards associated with the affected products. In addition, this updated notification was also informing their customers, if they have any patients with the affected product implanted, the Recalling Firm's investigation did not reveal any reason to suggest that there is any imminent danger to the patient since the noted failure was primarily due to a fall from height. It is up to the medical discretion of the treating physicians regarding whether or not to notify the patient. It is also up to the medical discretion of the treating physician whether or not to consider a revision surgery to replace the affected segment. Any questions, contact: -Director of Customer Service at 949-328-3366, or -Vice President of
Initiated
2021-08-27
Root cause
Device Design
Terminated
2023-01-20
Distribution
U.S. Nationwide: CA, CO, and NV No O.U.S.
Recall number
Z-0110-2022
Recall status
Terminated
CDRH record ID
189301
FDA record created
2021-10-14
Recall event number
88692

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗