Open original FDA record ↗Recall record
- FDA identifier
- Z-0110-2022
- Company in FDA record
- Uoc Usa Inc
- FDA record creation date
- Oct 14, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to product not meeting specification for concentricity and failing to meet requirements of dynamic fatigue testing.
Source facts
- Action
- On 08/27/2021, the Recalling Firm communicated via email a letter to customers/US surgeons informing them that they are initiating a product recall limited to its 25 mm segment from its knee system due to an out of specification condition that has the potential to pose a risk to patients. Customers are instructed to return unused products. For questions, contact the Director of Customer Service at 949-328-3366 or the Vice President of Marketing at 949-501-8575. On 09/16/2021, the firm sent an "URGENT: MEDICAL DEVICE RECALL" notification letter to additional customers, who have not used or received any of the affected product, just to make them aware of the recall. The Urgent Medical Device Recall notification identified the reason for the recall, the potential risks associate with the use of the affected product, and instructed these customers that to notify the recalling firm is they have any of the affected products. The Recalling Firm can be contacted via email harsh.gupta@unitedorthopedic.com or via phone at 949-501-8575. On 09/24/2021, the Recalling Firm communicated an updated "URGENT: MEDICAL DEVICE RECALL" Notification Letter informing customers that had received the affected recalled devices, of the recall action and the potential hazards associated with the affected products. In addition, this updated notification was also informing their customers, if they have any patients with the affected product implanted, the Recalling Firm's investigation did not reveal any reason to suggest that there is any imminent danger to the patient since the noted failure was primarily due to a fall from height. It is up to the medical discretion of the treating physicians regarding whether or not to notify the patient. It is also up to the medical discretion of the treating physician whether or not to consider a revision surgery to replace the affected segment. Any questions, contact: -Director of Customer Service at 949-328-3366, or -Vice President of
- Initiated
- 2021-08-27
- Root cause
- Device Design
- Terminated
- 2023-01-20
- Distribution
- U.S. Nationwide: CA, CO, and NV No O.U.S.
- Recall number
- Z-0110-2022
- Recall status
- Terminated
- CDRH record ID
- 189301
- FDA record created
- 2021-10-14
- Recall event number
- 88692