Open original FDA record ↗Recall record
- FDA identifier
- Z-0111-06
- Company in FDA record
- Ekos Corp
- FDA classification date
- Nov 1, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
During two separate procedures, the user experienced difficulty placing the EKOS ultrasound core (USC) into the EKOS drug delivery catheter (DDC) when transiting tight bend radii; specifically the aortic bifurcation. In one case, the USC fractured within the DDC when the user attempted to withdraw the USC after resistance was encountered
Source facts
- Action
- On 11/22/04 the firm issued letters to those customers that had product. The letters advised of the problem and provided amended use instructions.The firm continued to ship product and on 3/10/05 the firm visited consignees, replacing the catheters.
- Initiated
- 2004-11-22
- Root cause
- Other
- Terminated
- 2005-11-09
- Distribution
- Kits were distributed to 6 hospitals in the US, 1 each located in AZ, OH, WA, FL, IL, and CA. The firm also distributed to a customer in Germany.
- Recall number
- Z-0111-06
- Recall status
- Terminated
- CDRH record ID
- 41805
- FDA classified
- 2005-11-01
- Recall event number
- 33402