Z-0111-06: Lysus Infusion System - Lysus Drug Delivery Catheter: The Lysus Infusion System employs high frequency

Recall record Z-0111-06 for Ekos Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0111-06
Company in FDA record
Ekos Corp
FDA classification date
Nov 1, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

During two separate procedures, the user experienced difficulty placing the EKOS ultrasound core (USC) into the EKOS drug delivery catheter (DDC) when transiting tight bend radii; specifically the aortic bifurcation. In one case, the USC fractured within the DDC when the user attempted to withdraw the USC after resistance was encountered

Source facts

Action
On 11/22/04 the firm issued letters to those customers that had product. The letters advised of the problem and provided amended use instructions.The firm continued to ship product and on 3/10/05 the firm visited consignees, replacing the catheters.
Initiated
2004-11-22
Root cause
Other
Terminated
2005-11-09
Distribution
Kits were distributed to 6 hospitals in the US, 1 each located in AZ, OH, WA, FL, IL, and CA. The firm also distributed to a customer in Germany.
Recall number
Z-0111-06
Recall status
Terminated
CDRH record ID
41805
FDA classified
2005-11-01
Recall event number
33402

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗