Z-0111-2024: Astral 100 and Astral 150 ventilators

Recall record Z-0111-2024 for ResMed Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0111-2024
Company in FDA record
ResMed Ltd.
FDA classification date
Oct 23, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

If ventilator is on internal battery, not intended to serve as a primary power source, low/critically low battery alarms will sound, but a fault leads to sudden power loss. If power fails, then ventilation stops and a Total Power Failure (TPF) alarm should sound, but it's powered by a supercapacitor, which degrades over time, which may cause TPF alarm to sound for less than 2 minutes or not at all

Source facts

Action
On 9/13/23 Field Safety Notices were emailed to distributors, institutions, medical facilities. The notices states to immediately provide a copy of the notice and the patient/carer letter to all patients, carers, and relevant healthcare professionals. The notices states: Continue using your Astral ventilator, provided you continue to follow the instructions detailed in the Astral User Guide. -Connect the ventilator to mains power whenever possible. In the event of battery failure, connect to mains power immediately to resume ventilation. - Internal battery is not intended to serve as a primary power source. It should only be used when other sources are not available or briefly when necessary. -External power source should always be available for ventilator-dependent patients. -External power source should be in use in mobile situations, including when mains power is unavailable or disrupted. Do not rely solely on the internal battery for mobile use. - Ensure the external battery is sufficiently charged before using in mobile situations. - For ventilator-dependent patients, always have alternate ventilation equipment available, such as a back-up ventilator, manual resuscitator or similar device. - Ventilator-dependent patients should be continuously monitored by qualified personnel or adequately trained carers. Consistently follow the 2-year ventilator preventative maintenance schedule to help detect and replace circuit boards with degraded supercapacitors. A second phase will be communicated to customers. Follow the updated service process detailed in Tech Note 1063673; not sent to patients/carers. Complete/return the acknowledgment form: astral-postmarket@resmed.com. Questions - contact your local ResMed contact. On 3/18/24, a second Field Safety Notice was sent requesting that devices be returned to a Service Center as soon as possible, but no later than the end of March 2025 to receive software update R6.2 (SX544-0603).
Initiated
2023-09-13
Root cause
Device Design
Distribution
Worldwide - US Nationwide distribution including in the states of CA, NJ, VA, IN, MI, WV, NY, WI, ND, IL, NC, FL, CO, TX, AL, LA, MD, OR, AZ, MN, WA, PA, SC, MA, ID, GA, SD, TN, NH, OH, MS, CT, IA, KY, UT, NE, MO, NV, AR, WY, RI, KS, VT, ME, OK, PR, MO, DC and the countries of Taiwan, Israel, Spain, Slovenia, Belgium, Italy, Australia, Guatemala, Argentina, Uruguay, Chile, France, Martinique, Germany, French Polynesia, Korea, Egypt, Colombia, United Kingdom, Jersey, Ecuador, Reunion, Greece, Brazil, Cyprus, Saudi Arabia, Czech Republic, Thailand, New Caledonia, Mauritius, Canada, Vietnam, Indonesia, Pakistan, Japan, Guadeloupe, South Africa, Iran, Hungary, Tunisia, Nepal, Iceland, New Zealand, Austria, Portugal, Malaysia, Norway, Netherlands, Sweden, Lebanon, Finland, Singapore, United Arab Emirates, Peru, Oman, Philippines, Bahrain, French Guiana, Mexico, Sri Lanka, India, Kuwait, Bangladesh, Kenya, Myanmar, Turkey, Bulgaria, Croatia, Hong Kong, Saint Martin, Qatar, Jordan, Romania, Maldives, Denmark, Poland.
Recall number
Z-0111-2024
Recall status
Open, Classified
CDRH record ID
203491
FDA classified
2023-10-23
Recall event number
93122

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗