Z-0112-2015: Covidien Puritan Bennett 980 Ventilator, Rx ONLY. Suitable for service in a hospital (institutions) an

Recall record Z-0112-2015 for Nellcor Puritan Bennett Inc. (dba Covidien LP). View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0112-2015
Company in FDA record
Nellcor Puritan Bennett Inc. (dba Covidien LP)
FDA classification date
Oct 28, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

A software issue may lead to ventilator inoperative situations.

Source facts

Action
The firm sent Urgent: Field Correction Action letters, dated October 3, 2014, to consignees. The letter identified the affected product and the reason for the recall. Customers were provided actions to be taken, as well as, important safety reminders. The letter informed customers that they may continue to use their ventilators pending the software correction as long as two gas sources are connected to the ventilator at all times. Customers are to complete the attached acknowledgement and receipt form and fax it to Covidien at the number provided. For further assistance, customers are to contact the Technical Support Department at 1-800-255-6774.
Initiated
2014-10-01
Root cause
Software design
Terminated
2016-05-02
Distribution
Worldwide Distribution -- USA, including the states of CA, CO, FL, GA, KY, MA, MN, NC, NY, OH, OK, PA, SC, TN, TX, UT, WI; and, the countries of Canada, Mexico, Saudi Arabia, South Africa, and United Arab Emirates.
Recall number
Z-0112-2015
Recall status
Terminated
CDRH record ID
130226
FDA classified
2014-10-28
Recall event number
69320

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗