Skip to content

Public FDA record

Z-0112-2024: Monoject(TM) 1 mL Tuberculin Syringe Luer-Lock Tip Soft Pack, Model Number/Product Code: 1180100777

Recall record Z-0112-2024 for Cardinal Health 200, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0112-2024
Company in FDA record
Cardinal Health 200, LLC
FDA classification date
Oct 13, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Syringe contains conical tip that is not compatible with certain needleless IV connector(s).

Source facts

Action
Cardinal Health notified consignees on about 09/13/2023 via letter titled "URGENT MEDICAL DEVICE PRODUCT CORRECTION." Customers were instructed to review inventory for the affected product code and lot, ensure all personnel are informed cease utilizing the affected product with needleless IV connectors, post a copy of the notification in the storeroom where the product is stored, and notify any customers if the product has been further distributed/forwarded. Additionally, customers were asked to complete and return the included acknowledgment form.
Initiated
2023-09-13
Root cause
Process change control
Distribution
Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY
Recall number
Z-0112-2024
Recall status
Open, Classified
CDRH record ID
203290
FDA classified
2023-10-13
Recall event number
93019

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗