When a spectrum of a Pulsed Wave Doppler shows an angle correction unequal to 0 is reloaded from the archive or the image is sent to a DICOM station, this angle correction is mistakenly disregarded resulting in an underestimation of the flow velocities.
Source facts
Action
GE Healthcare sent customer letters on July 15, 2007 to affected customers. The letter included a short term mitigation recommendation and a long-term correction. The long-term correction will be software upgrade provided free of charge to all affected customers.
Initiated
2007-07-06
Root cause
Software design
Terminated
2008-10-21
Distribution
Nationwide.
Recall number
Z-0113-2008
Recall status
Terminated
CDRH record ID
55383
FDA classified
2007-11-07
Recall event number
39400