Open original FDA record ↗Recall record
- FDA identifier
- Z-0114-2013
- Company in FDA record
- GE Healthcare, LLC
- FDA classification date
- Oct 17, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
It was discovered of a potential failure of the spectral filter mechanism inside the collimator of the innova 3100, innova 3100-IQ and innova 3131-IQ systems.
Source facts
- Action
- GE Healthcare will, without charge, remedy the defect or bring the product into compliance with each applicable Federal standard in accordance with a plan to be approved by the Secretary of Health and Human Services. A GE Healthcare Service Representative will replace the collimator to address the issue. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. Further questions please call 262-513-4122.
- Initiated
- 2011-12-20
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2013-09-18
- Distribution
- Nationwide distribution
- Recall number
- Z-0114-2013
- Recall status
- Terminated
- CDRH record ID
- 112314
- FDA classified
- 2012-10-17
- Recall event number
- 63027