Open original FDA record ↗Recall record
- FDA identifier
- Z-0115-2007
- Company in FDA record
- Radiometer America Inc
- FDA classification date
- Oct 28, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Incorrect FI02 values: programming issue can cause incorrect values to be transmitted to the LIS when : 1)- The FIO2 result is edited in the manual sample processing mode--2)- An existing result is opened and FI02 is then edited and sent.....In both cases RADIANCE will transmit the original FI02 value, not the value that was edited.
Source facts
- Action
- The recalling firm initially contacted their customers and informed them of the software problem and the upcoming field correction for the software, via telephone on 7/6/2006. A follow-up letter dated 7/21/2006, was also sent to the customers.
- Initiated
- 2006-07-06
- Root cause
- Other
- Terminated
- 2011-11-18
- Distribution
- Worldwide-The six USA units (equpped with Radiance software) were distributed to a total of two (2) U.S. customers: University of Missouri Hospitals and Clinics in Columbia, MO (4 units) and Vail Valley Medical Center, Vail, CO (2 units). World wide accounts pending submission. Three (3) Canadian customers. They are: 1) Seven Oaks General Hospital MB Canada; and 2) St. Mary''s General Hospital, 911 Queens Blvd, Ontario Canada; and 3) University of Alberta Hospital, 84440 112th Street, Alberta, Canada.
- Recall number
- Z-0115-2007
- Recall status
- Terminated
- CDRH record ID
- 48877
- FDA classified
- 2006-10-28
- Recall event number
- 36539