Open original FDA record ↗Recall record
- FDA identifier
- Z-0115-2009
- Company in FDA record
- Stryker Howmedica Osteonics Corp.
- FDA classification date
- Oct 20, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Stryker Orthopaedics became aware that there is a potential inability to release the Scorpio Series 7000 Tibial Impactor/Extractor from a tibial baseplate during use.
Source facts
- Action
- US recall notification (product correction) letters were send out by Federal Express with return receipt on July 2 and 9, 2008. Advisory notices were sent to foreign customers on July 2, 2008. Both letters include a Product Correction Bulletin that instructs all used to lubricate the divide with an approved medical lubricant, especially the extractor shaft and threads. It is recommended that all moving parts be evaluated for functionality and useful service prior to each surgery and instructions on how to perform this process. Contact Stryker Howmedica Osteonics Corp. at 1-201-831-5918 for assistance.
- Initiated
- 2008-07-02
- Root cause
- Component design/selection
- Terminated
- 2012-08-17
- Distribution
- Worldwide including the US and Asia Pacific, Canada, Europe, Middle East, Africa, Japan and Latin America.
- Recall number
- Z-0115-2009
- Recall status
- Terminated
- CDRH record ID
- 72482
- FDA classified
- 2008-10-20
- Recall event number
- 48995