Z-0115-2016: MRIdian ViewRay Radiation Therapy System, ViewRay Is indicated for stereotactic radiosurgery and precis

Recall record Z-0115-2016 for Viewray Incorporated. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0115-2016
Company in FDA record
Viewray Incorporated
FDA classification date
Oct 9, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

ViewRay discovered that in the event that an encoder breaks or fails on the Patient Handling System (PHS, or couch), when attempting to restart the system, the couch could move unexpectedly.

Source facts

Action
ViewRay sent an Customer Advisory Notice dated August 27, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. It is recommended that if you have a patient on the couch and you receive the fault ''The patient table has experienced an unintended movement. This may be due to the patient or operator leaning on the table. Clear the fault and retry. If the problem persists call Service" on the TPDS,first attempt to clear the fault. If it does not clear and requires you to restart the couch software (same switch as restarting the RTCS) you should manually remove the patient from the couch before restarting the system. A software patch to address this issue is planned for early 2016. In the meantime please practice our recommendation of removing the patient from the couch prior to restarting the RTCS. If you have any questions or concerns regarding this issue, please do not hesitate to contact ViewRay Customer Support at support@viewray.com or call the ViewRay Customer Support center at 855-286-8875.
Initiated
2015-08-27
Root cause
Device Design
Terminated
2017-06-09
Distribution
US Distribution to the states of : CA, MO and WI., and Internationally to Korea.
Recall number
Z-0115-2016
Recall status
Terminated
CDRH record ID
140076
FDA classified
2015-10-09
Recall event number
72166

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗