Z-0115-2021: Atellica IM HER-2/neu (H2n) assay, 50 Test Kit, Catalog No. SMN 10995591, UDI Codes: (01)00630414599151(1

Recall record Z-0115-2021 for Siemens Healthcare Diagnostics, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0115-2021
Company in FDA record
Siemens Healthcare Diagnostics, Inc
FDA record creation date
Oct 16, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Siemens confirmed that the Upper Limit of Normal (ULN) as claimed in the Instructions for Use (IFU) was no longer achieved. This can potentially result in false positive or false negative results.

Source facts

Action
All affected customers were sent a letter on 08/25/2020 stating the following: Siemens Healthcare Diagnostics Inc. (Siemens Healthineers) notified customers on August 25th, 2020. Letter states reason for recall, health risk and action to take. Actions to be taken by their customers include: - Reviewing the letter with their Medical Director. - Customers may continue use of the product lots referenced in Table 1 until they receive replacement product in their laboratory. Refer to the Table 3 for product biasinformation. - If you are currently using the products listed in Table1, review your inventory of these products, as well as the associated product QC and MCM, and order replacement products by completing the Field Correction Effectiveness Check Form attached to the letter. - Upon acceptance of the replacement lots referenced in Table 2 New Lot Combinations column, discontinue use of and discard the products listed in Table 2 Current Lot Combinations column. Refer to Figures 1 through 4 for expected results with replacement lots. - Complete and return the Field Correction Effectiveness Check Form attached to this letter. Retain the letter with your laboratory records and forward the letter to those who may have received this product.
Initiated
2020-08-25
Root cause
Process control
Terminated
2023-03-10
Distribution
US Nationwide distribution.
Recall number
Z-0115-2021
Recall status
Terminated
CDRH record ID
183649
FDA record created
2020-10-16
Recall event number
86383

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗