Z-0115-2022: Elekta Medical Linear Accelerator (EMLA): Elekta Harmony Pro, REF 1555143; Elekta Infinity, REF 1552091,

Recall record Z-0115-2022 for ELEKTA SOLUTIONS AB. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0115-2022
Company in FDA record
ELEKTA SOLUTIONS AB
FDA record creation date
Oct 14, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected matter.

Source facts

Action
The recalling firm issued an Urgent Important Field Safety Notification dated September 2021 via email on 9/28/2021. The Subject of the letter was addressed for the service user regarding precautions to be taken when handling diode D1. The notification described the details of the failure and that Elekta is currently working on a solution to eliminate this failure mode. The notice was to be placed in an area accessible to all users until the correction is made. Appropriate personnel who work with this product are to be advised on the content of the letter. Corrective actions were provided regarding removing the modulator covers and the Personal Protective Equipment that must be worn by the engineer, including a safety visor, protective jacket, and gloves. Instructions for removing the RDO diode (D1) were provided. An Acknowledgment Form was included for completion and return to Elekta no later than 30 days for acknowledgement that the service user has read and understands the notification and accepts the implementation of any given recommendation.
Initiated
2021-09-28
Root cause
Nonconforming Material/Component
Terminated
2024-06-14
Distribution
Distribution was to ND and NC. There was no military/government distribution.
Recall number
Z-0115-2022
Recall status
Terminated
CDRH record ID
189498
FDA record created
2021-10-14
Recall event number
88771

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗