Open original FDA record ↗Recall record
- FDA identifier
- Z-0115-2022
- Company in FDA record
- ELEKTA SOLUTIONS AB
- FDA record creation date
- Oct 14, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected matter.
Source facts
- Action
- The recalling firm issued an Urgent Important Field Safety Notification dated September 2021 via email on 9/28/2021. The Subject of the letter was addressed for the service user regarding precautions to be taken when handling diode D1. The notification described the details of the failure and that Elekta is currently working on a solution to eliminate this failure mode. The notice was to be placed in an area accessible to all users until the correction is made. Appropriate personnel who work with this product are to be advised on the content of the letter. Corrective actions were provided regarding removing the modulator covers and the Personal Protective Equipment that must be worn by the engineer, including a safety visor, protective jacket, and gloves. Instructions for removing the RDO diode (D1) were provided. An Acknowledgment Form was included for completion and return to Elekta no later than 30 days for acknowledgement that the service user has read and understands the notification and accepts the implementation of any given recommendation.
- Initiated
- 2021-09-28
- Root cause
- Nonconforming Material/Component
- Terminated
- 2024-06-14
- Distribution
- Distribution was to ND and NC. There was no military/government distribution.
- Recall number
- Z-0115-2022
- Recall status
- Terminated
- CDRH record ID
- 189498
- FDA record created
- 2021-10-14
- Recall event number
- 88771