Open original FDA record ↗Recall record
- FDA identifier
- Z-0116-2009
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Oct 20, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Three issues: unexpected rotation of gantry, unexpected movement of table between beams during patient setup, and single exposure images will be overwritten by double exposure images.
Source facts
- Action
- A letter marked "Urgent" dated July 4, 2008 was sent to customers. The letter provided a detailed description of the issues, the conditions under which these issues may occur, as well as recommendations to avoid these issues. Customer notification TH022/08/S was released on July 11, 2008. A software fix is currently being validated. All consignees notified by certified mail. Contact Siemens Medical Solutions USA, Inc. at 1-925-246-8407 for assistance.
- Initiated
- 2008-07-04
- Root cause
- Software change control
- Terminated
- 2010-12-22
- Distribution
- Nationwide in LA, WI, DE, PA and OH, and internationally in Netherlands, Germany, Poland, Norway, Australia, Hungary, Belgium and France
- Recall number
- Z-0116-2009
- Recall status
- Terminated
- CDRH record ID
- 72541
- FDA classified
- 2008-10-20
- Recall event number
- 49024