Open original FDA record ↗Recall record
- FDA identifier
- Z-0117-2012
- Company in FDA record
- Philips Healthcare Inc.
- FDA classification date
- Oct 27, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Front plate broke in to two pieces causing the C-Arm to suspend on the single central bolt
Source facts
- Action
- Philips Healthcare Inc., sent an "URGENT - Field Safety Notice" dated September 30, 2011 to all affected customers. The notice identifies the following: what the problem is and under what circumstances it can occur; the actions that should be taken by the customer/user in order to prevent risks for patients or users and the actions planned by Philips Medical Systems to correct the problem. The notice contains important information for the continued safe and proper use of the equipment and customers were advised to retain a copy with the equipment Instruction for Use. The notice stated that Philips will replace the front plates for systems on the Unit Affected List by the end of June 2012. The notice instructed customers to contact Philips Healthcare representative for further information or support.
- Initiated
- 2011-10-04
- Root cause
- Nonconforming Material/Component
- Terminated
- 2017-01-12
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of: Australia, Austria, Belgium, Bharain, Canada, Croatia, Denmark, France, Germany, Greece, Italy, Japan, Netherlands, Norway, Portugal, Russian Fed., Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, and United Kingdom.
- Recall number
- Z-0117-2012
- Recall status
- Terminated
- CDRH record ID
- 104360
- FDA classified
- 2011-10-27
- Recall event number
- 60064