Skip to content

Public FDA record

Z-0117-2013: HiRes 90K devices with Helix Electrodes Model CI 1400-02H The HiResolution Bionic Ear System is a coch

Recall record Z-0117-2013 for Advanced Bionics Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0117-2013
Company in FDA record
Advanced Bionics Corporation
FDA classification date
Oct 22, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

Advanced Bionics is recalling the the HiRes 90K cochlear implant with HiFocus Helix Electrode because of loose-fitting insertion tool assemblies.

Source facts

Action
The firm, Advanced Bionics, sent a " RECALL NOTIFICATION" letter dated September 14, 2007 to all their customers who purchased the HiRes 90K devices with Helix Electrodes Model CI 1400-02H. The letter identified the product, problem, and actions to be taken. The customers were instructed to immediately locate and return the affected product to Advanced Bionics. The firm's Service representative will contact the customers within 48 hours to retrieve the affected product and send a replacement as needed. The customers were also instructed to complete and return the enclosed Notification Acknowledgement form via fax to: (661) 362-1511. If you have any questions regarding this letter, call Advanced Bionics at (877) 829-0026 between the hours of 5 AM and 5 PM Pacific Time, Monday through Friday.
Initiated
2007-09-14
Root cause
Nonconforming Material/Component
Terminated
2012-10-22
Distribution
Worldwide distribution: USA (nationwide) including state: IL and country of: Europe.
Recall number
Z-0117-2013
Recall status
Terminated
CDRH record ID
113560
FDA classified
2012-10-22
Recall event number
63353

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗