- FDA identifier
- Z-0117-2016
- Company in FDA record
- Natus Neurology Inc
- FDA classification date
- Oct 13, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
When using the NicoletOne v5.94 software, after exiting the impedance check function and returning to the EEG screen, the impedance check signal remains active in waveform, obscuring the EEG signals.
Source facts
- Action
- Natus Neurology sent an Ugent Medical Device Corrrection Notification letter dated September 9, 2015 to all affected consignees. The letter described the problem and the product involved in the recall. Consignees were informed consignees that a Field Service Technician will contact them to install the new software. Once the software is installed they will ask the Recipient or Field service Technician to complete and return the Verification Form. For questions regarding this recall call 608-829-8600.
- Initiated
- 2015-09-11
- Root cause
- Software change control
- Terminated
- 2016-03-08
- Distribution
- Worldwide distribution. US nationwide, INDIA, ROMANIA, DENMARK, JAPAN, AUSTRALIA, RUSSIA, BANGLADESH, SRI LANKA, BRUNEI, and UNITED ARAB EMIRATES.
- Recall number
- Z-0117-2016
- Recall status
- Terminated
- CDRH record ID
- 139574
- FDA classified
- 2015-10-13
- Recall event number
- 71991