Certain Aespire and Aestiva Anesthesia Systems were noted to have a vulnerability to a cyber-attack when connected to the hospital network. An insufficiently secured terminal server may provide an opportunity for a malicious actor that has already penetrated the hospital network to send fraudulent flow sensor correction parameters.
Source facts
Action
The firm sent an urgent medical device correction notice to their consignees by letter on 07/19/2019. The letter explained the issue and requested that the consignee ensure that the devices are connected to a sufficiently secure terminal server.
Initiated
2019-07-19
Root cause
Software in the Use Environment
Terminated
2026-04-13
Distribution
Worldwide
Recall number
Z-0117-2020
Recall status
Terminated
CDRH record ID
175186
FDA record created
2019-12-17
Recall event number
83425