Open original FDA record ↗Recall record
- FDA identifier
- Z-0118-2013
- Company in FDA record
- Elekta, Inc.
- FDA classification date
- Oct 23, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
During the planning process the beam weights can be changed by the customers. When the beam weights are changed and renumbered and this happens during the dose calculation it is possible to save and reopen the plan without any error messages.
Source facts
- Action
- Nucletron sent a Customer Information Bulletin 555.00248 to all affected customers. The bulletin identified the product, problem, and actions to be taken by the customers. The bulletin contained information on how an incorrect dose association can be prevented by not renumbering beams during recalculation. A Read Confirmation Form was included for customers to complete and return. Contact your local Nucleton Service Representative for questions regarding this notice.
- Initiated
- 2012-08-21
- Root cause
- Software design
- Terminated
- 2014-11-03
- Distribution
- Nationwide Distribution-including the states of AL, CA, DC, KS, LA, MO, NB, NY, NC, ND, OH, OR, PA, RI, SC, TX, and WA.
- Recall number
- Z-0118-2013
- Recall status
- Terminated
- CDRH record ID
- 113697
- FDA classified
- 2012-10-23
- Recall event number
- 63405