Z-0118-2019: AS-OBGYN Information System version 7.824.x

Recall record Z-0118-2019 for AS Software, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0118-2019
Company in FDA record
AS Software, Inc.
FDA record creation date
Oct 16, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Any change made to current ultrasound machine software or hardware upgrades, adding or modifying custom fields, or adding a new Ultrasound Machine, must be tested prior to the use in production environment to ensure that ultrasound measurements are crossing over correctly to the AS system

Source facts

Action
AS Software issued letter by email on 9/20/18 to notify customers that any change made to their ultrasound machine such as software or hardware upgrades, adding or modifying custom fields, or adding a new Ultrasound Machine, are tested prior to the use in the production environment. Medical Device Correction Return Response form to be signed and returned. The AS Software Support Department can be reached Monday through Friday, 9:00 AM to 6:00 PM EST via phone: 201-541-1900 option 2 or e-mail: info@as-software.com
Initiated
2018-09-20
Root cause
Device Design
Terminated
2019-04-19
Distribution
US Nationwide.
Recall number
Z-0118-2019
Recall status
Terminated
CDRH record ID
167364
FDA record created
2018-10-16
Recall event number
81057

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗