Open original FDA record ↗Recall record
- FDA identifier
- Z-0118-2019
- Company in FDA record
- AS Software, Inc.
- FDA record creation date
- Oct 16, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Any change made to current ultrasound machine software or hardware upgrades, adding or modifying custom fields, or adding a new Ultrasound Machine, must be tested prior to the use in production environment to ensure that ultrasound measurements are crossing over correctly to the AS system
Source facts
- Action
- AS Software issued letter by email on 9/20/18 to notify customers that any change made to their ultrasound machine such as software or hardware upgrades, adding or modifying custom fields, or adding a new Ultrasound Machine, are tested prior to the use in the production environment. Medical Device Correction Return Response form to be signed and returned. The AS Software Support Department can be reached Monday through Friday, 9:00 AM to 6:00 PM EST via phone: 201-541-1900 option 2 or e-mail: info@as-software.com
- Initiated
- 2018-09-20
- Root cause
- Device Design
- Terminated
- 2019-04-19
- Distribution
- US Nationwide.
- Recall number
- Z-0118-2019
- Recall status
- Terminated
- CDRH record ID
- 167364
- FDA record created
- 2018-10-16
- Recall event number
- 81057