Z-0118-2022: ROSA One 3.1 Brain application The device is intended for the spatial positioning and orientation of ins

Recall record Z-0118-2022 for MEDTECH SAS. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0118-2022
Company in FDA record
MEDTECH SAS
FDA record creation date
Oct 25, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The firm has become aware of a software anomaly affecting ROSA One 3.1 Brain application which led to the inaccurate placement of an electrode during surgery. The firm has received 3 global complaints related to the issue. An incorrect trajectory could result in serious injury or death if undetected during surgery.

Source facts

Action
On September 22, 2021, the firm, Zimmer Biomet, issued "URGENT MEDICAL DEVICE CORRECTION" letters to affected consignees. Customers were informed of steps that may result in the issue and that should be avoided: After the registration of the patient during the ROSA One 3.1 Brain application procedure, the issue may occur when the device shuts down either manually or unexpectedly. After reboot, if the trajectory is interrupted by deactivating the pedal, the device gives the user an option to Return to HOME , Clear the robotic arm , or Continue . If the user chooses to Clear the robotic arm , the device prompts the user to perform an empty device calibration. If the user performs this step and presses start, the issue will occur, which results in the device driving to an incorrect trajectory. Pending the implementation of a corrective action, Medtech S.A - Zimmer Biomet - advises that the actions detailed below to be taken without delay by all users of the ROSA One 3.1 Brain application device. Required actions by users to prevent incorrect instrument configuration (Workaround): If a shutdown occurs during guidance mode, any empty calibration window that would subsequently appear during the session should be closed by pressing UNDO. Customers were provided with a laminated warning label to attach to the robot. A Zimmer Biomet engineer will visit each customer site to implement a new software version to correct the issue once it is available. It is estimated that the updates will begin approximately in February 2022 and be completed by May 2022. If you have further questions or concerns, please contact your Zimmer Biomet representative or call the Product Surveillance Manager at stephanie.leppo@zimmerbiomet.com.
Initiated
2021-09-22
Root cause
Software design
Distribution
US Nationwide distribution and International Distribution to countries of: Australia, Taiwan, China, Thailand, Japan, India, South Korea, and the Netherlands.
Recall number
Z-0118-2022
Recall status
Open, Classified
CDRH record ID
189497
FDA record created
2021-10-25
Recall event number
88746

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗