Skip to content

Public FDA record

Z-0118-2023: AutoPulse Li-ion Battery used on the AutoPulse Resuscitation System Model 100. used as an adjunct to man

Recall record Z-0118-2023 for ZOLL Circulation, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0118-2023
Company in FDA record
ZOLL Circulation, Inc.
FDA classification date
Oct 21, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to increase in Li-Ion Battery failures

Source facts

Action
On 09/22/2022, the firm sent an "MEDICAL DEVICE PRODUCT CORRECTION NOTICE" via First Class Mail to customers informing them that ZOLL Circulation observed an increase in battery complaints of fully charged batteries no powering on the AutoPulse Platform. Investigations revealed that batteries stored at elevated temperatures were failing and are potentially due to a supplier switching to an alternate electrical component and thermal variation within specification of a second electrical component. Customers are instructed to: 1. Complete the Customer Acknowledgement/No-Charge Battery Replacement Request Form and email it to APBattery2022@zoll.com as soon as possible. 2. While waiting for the replacement batteries to arrive, do not store the affected batteries in a hot environment prior to use such as in a hot vehicle, in the sun, or in a hot room. Storing the affected batteries in a hot environment may prevent the batteries from powering on the AutoPulse Platform. Note that the affected batteries will power on the AutoPulse Platform after they are allowed to cool.
Initiated
2022-09-22
Root cause
Device Design
Distribution
U.S.: AK, AL, AR, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, RN, TX, UT, VA, VT, Virgin Islands, WA, WI, WV and WY. O.U.S.: United Arab Emirates, Austria, Australia, Azerbaijan, Brazil, Belarus, Canada, Switzerland, China, Costa Rica, Germany, Denmark, Egypt, France, United Kingdom, Greece, Hong Kong, Croatia, Hungary, Israel, India, Iraq, Italy, Jordan, Japan, South Korea, Moldova, Mexico, Malaysia, Netherlands, Norway, Panama, Philippines, Poland, Qatar, Romania, Russian, Saudi Arabia, Singapore, Slovenia, Taiwan, and Vietnam.
Recall number
Z-0118-2023
Recall status
Open, Classified
CDRH record ID
196113
FDA classified
2022-10-21
Recall event number
90879

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗