Open original FDA record ↗Recall record
- FDA identifier
- Z-0118-2023
- Company in FDA record
- ZOLL Circulation, Inc.
- FDA classification date
- Oct 21, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to increase in Li-Ion Battery failures
Source facts
- Action
- On 09/22/2022, the firm sent an "MEDICAL DEVICE PRODUCT CORRECTION NOTICE" via First Class Mail to customers informing them that ZOLL Circulation observed an increase in battery complaints of fully charged batteries no powering on the AutoPulse Platform. Investigations revealed that batteries stored at elevated temperatures were failing and are potentially due to a supplier switching to an alternate electrical component and thermal variation within specification of a second electrical component. Customers are instructed to: 1. Complete the Customer Acknowledgement/No-Charge Battery Replacement Request Form and email it to APBattery2022@zoll.com as soon as possible. 2. While waiting for the replacement batteries to arrive, do not store the affected batteries in a hot environment prior to use such as in a hot vehicle, in the sun, or in a hot room. Storing the affected batteries in a hot environment may prevent the batteries from powering on the AutoPulse Platform. Note that the affected batteries will power on the AutoPulse Platform after they are allowed to cool.
- Initiated
- 2022-09-22
- Root cause
- Device Design
- Distribution
- U.S.: AK, AL, AR, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, RN, TX, UT, VA, VT, Virgin Islands, WA, WI, WV and WY. O.U.S.: United Arab Emirates, Austria, Australia, Azerbaijan, Brazil, Belarus, Canada, Switzerland, China, Costa Rica, Germany, Denmark, Egypt, France, United Kingdom, Greece, Hong Kong, Croatia, Hungary, Israel, India, Iraq, Italy, Jordan, Japan, South Korea, Moldova, Mexico, Malaysia, Netherlands, Norway, Panama, Philippines, Poland, Qatar, Romania, Russian, Saudi Arabia, Singapore, Slovenia, Taiwan, and Vietnam.
- Recall number
- Z-0118-2023
- Recall status
- Open, Classified
- CDRH record ID
- 196113
- FDA classified
- 2022-10-21
- Recall event number
- 90879