Z-0118-2025: IMMAGE Immunochemistry Systems IGM Immunoglobulin M Reagent REF 447610 IGM reagent, when used in conjunc

Recall record Z-0118-2025 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0118-2025
Company in FDA record
Beckman Coulter Inc.
FDA classification date
Oct 17, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Firm identified two lots of reagents that are not meeting the sensitivity statement in the IFU

Source facts

Action
During the week of 08/15/2024, the firm an "URGENT MEDICAL DEVICE RECALL" Letter via first class mail and/or email. The letter informed customer that Beckman Coulter discovered that IMMAGE Immunochemistry Systems IGM Immunoglobulin M reagent lots are not meeting the sensitivity statement in the IFU: Sensitivity is defined as the lowest measurable concentration which can be distinguished from zero with 95% confidence. Sensitivity for IGM determination is 4.2 mg/dL. . Customers are instructed to: -Discontinue use of affected IMMAGE Immunochemistry Systems IGM Immunoglobulin M reagent lots and to follow product replacement instructions. -Review the Safety Data Sheet (SDS) and discard according to their laboratory's Standard Operating Procedures and/or their local regulations. SDS is available on www.beckmancoulter.com -Share the content of Recall Letter with their lab and/or medical director regarding the need to perform a retrospective review of patient results. Beckman Coulter is no longer distributing the affected lots. For questions, contact Customer Support Center: . From website: http://www.beckmancoulter.com . By phone: call 1-800-854-3633 in the United States. o Outside the United States, contact local Beckman Coulter representative. . For customers in the United States, if need replacement product: o Complete the attached Replacement Order Form and email to askbeckman@beckman.com or fax to (866) 294-7850 OR o Call Client Services at (800) 526-3821 . For customers in other geographies, contact local Beckman Coulter Representative for replacement.
Initiated
2024-07-31
Root cause
Employee error
Distribution
U.S.: AZ, CA, CO, FL, GA, HI, IL, MA, MI, MO, ND, NY, OH, PA, and TX O.U.S.: Australia, Belgium, Brazil, Chile, China, Czechia, Ecuador, France, Germany, Greece, Hong Kong, India, Italy, Kuwait, Mexico, Philippines, Poland, Portugal, Saudi Arabia, Slovakia, Spain, Sweden, Taiwan Province of China, and Turkey,
Recall number
Z-0118-2025
Recall status
Open, Classified
CDRH record ID
209901
FDA classified
2024-10-17
Recall event number
95285

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗