Open original FDA record ↗Recall record
- FDA identifier
- Z-0119-2008
- Company in FDA record
- Opti Medical Systems Inc
- FDA classification date
- Nov 9, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Incorrect results/stability failure; Actual device stability studies show sodium and chloride sensors age differently than anticipated when stored at the high end of labeled storage, therefore subsequent measurements give results outside the allowed error allowance before the 6 month expiration date shown on the device.
Source facts
- Action
- An Urgent Medical Device Recall Letter dated 16 July 2007, was issued via E mail and Fax on/or about 7/18/2007 directing consignees to remove unused cassettes from inventory and destroy them. Replacement cassettes will be shipped upon receipt of information regarding destruction by calling customer service at 800-490-6784. Sub-recall by distributors was requested.
- Initiated
- 2007-07-18
- Root cause
- Device Design
- Terminated
- 2007-12-13
- Distribution
- Worldwide Distribution-USA, Australia, Chile, China, Egypt, France, Great Britain, Guatemala, Indonesia, Italy, Korea, Lebanon, Mauritania, Morocco, Nepal, Russia, South Africa, Sweden, Tanzania, Thailand, and Venezuela.
- Recall number
- Z-0119-2008
- Recall status
- Terminated
- CDRH record ID
- 64467
- FDA classified
- 2007-11-09
- Recall event number
- 44880