Open original FDA record ↗Recall record
- FDA identifier
- Z-0119-2016
- Company in FDA record
- Zimmer Gmbh
- FDA classification date
- Oct 14, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
In about 5% of tested samples, the outer tray was found to be compromised. Risks: A damaged outer tray detected during surgery may cause a delay in surgery to obtain another implant. If a compromised packaging is not detected prior or during a surgery by the user, it may lead to periprosthetic infection. This may result in revision or multi-stage revision to treat the infection.
Source facts
- Action
- On 9/3/2015, URGENT MEDICAL DEVICE RECALL notifications were sent via courier to the affected distributors and Hospital Risk Managers with instructions for locating and inspecting the affected devices and contacting Zimmer if damage is observed. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. If after reviewing this notification you have questions concerning this notification, please call 1-877-946-2761.
- Initiated
- 2015-09-03
- Root cause
- Package design/selection
- Terminated
- 2016-03-15
- Distribution
- US Nationwide distribution.
- Recall number
- Z-0119-2016
- Recall status
- Terminated
- CDRH record ID
- 139914
- FDA classified
- 2015-10-14
- Recall event number
- 72123