Z-0119-2021: Olympus CHF-CB30S Choledoscope Intended Use: The device was designed to be used with an Olympus TV came

Recall record Z-0119-2021 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0119-2021
Company in FDA record
Olympus Corporation of the Americas
FDA record creation date
Oct 16, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The CHF-CB30S Device did not have a 510(k) clearance. Additionally, the device has been associated with breakage of the insertion section and protrusion of metal parts, as well as breakage and displacement of the rubber on the bending section during surgical procedures.

Source facts

Action
All affected customers were sent a "URGENT: MEDICAL DEVICE REMOVAL ACTION" letter on 08/31/2020 stating the following: Olympus Corporation notified direct customer accounts on 08/31/2020 via recall notification letter. Letter states reason for recall, health risk and action to take. Actions to be taken by customers including the following: 1. Inspect your inventory and identify any CHF-CB30S devices. 2. Cease any further use of any CHF-CB30S device you have, remove it from your inventory and quarantine it until it is shipped back to us. 3. Call your Olympus customer service representative at 1-888-524-7266 option 1. Olympus will issue a Return Material Authorization for you to return any CHF-CB30S at no charge. 4. Olympus will reimburse you for each returned CHF-CB30S. Please contact our customer Service representative as indicated in step 3 for the reimbursement amount for your CHF-CB30S choledoschope. 5. Access the Olympus recall portal to indicate that you have received this notification. go to https://olympusamerica.com/recall. Enter the recall number Recall-0365 and provide your contact information as indicated in the portal. 6. If you may have further distributed the CHF-CB30S, please identify your customers, notify them at once of this product recall, and appropriately document your notification process. Your notification to your customers may be enhanced by including a copy of this recall notification letter. OMSC requests you to report any patient injuries associated with any OMSC endoscope. Call our Technical Assistance Center (TAC) at 1-800-848-9024, option 1 to report complaints. If you have any questions or concerns, please do not hesitate to contact me directly at 484-896-5688 or at laura.storms@olympus.com.
Initiated
2020-08-31
Root cause
Device Design
Terminated
2022-12-06
Distribution
US Nationwide Distribution in states of: AK, AZ, CA, FL, GA, IL, LA, MA, ME, MI, MN, MS, NC, NM, NJ, NY, OR, SC, TN, WA, and WI; and International to: Japan.
Recall number
Z-0119-2021
Recall status
Terminated
CDRH record ID
183489
FDA record created
2020-10-16
Recall event number
86367

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗