Open original FDA record ↗Recall record
- FDA identifier
- Z-0119-2022
- Company in FDA record
- Windstone Medical Packaging, Inc.
- FDA record creation date
- Oct 29, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Monoject 0.9% Sodium Chloride Flush Syringe Manufactured by Cardinal Health and placed into kits. Potential for the plunger to draw back after air has been expelled and reintroduce air back into the syringe.
Source facts
- Action
- On 08/19/2021, recall notices were emailed to customers. Customers were asked to do the following: Identify remaining affected kits and follow the over-labeling instructions. The sticker should be placed in such a way that it would be immediately noticed but should not cover up or conceal any pertinent information such as Pack Number, Pack Name, Lot Number or Expiration Date. Sticker states in part: A recall of the prefilled syringe contained in this kit was announced. All other kit components are not affected. 1. At the time the kit is opened for use any prefilled syringe manufactured by Cardinal Health should be identified and set aside 2. The recalled syringe should be rendered unusable to protect against inadvertent use and disposed of pursuant to the medical waste policies in effect at your institution. Complete and return the reply form. If product was further distributed, notify your customers of the product recall Customers with questions or concerns can contact the recalling firm at 407-638-9924, www.alignedmedicalsolutions.com There is a recall notice on the website: http://www.alignedmedicalsolutions.com/ams_recall_08192021.htm
- Initiated
- 2021-08-19
- Root cause
- Device Design
- Distribution
- FL and PA
- Recall number
- Z-0119-2022
- Recall status
- Open, Classified
- CDRH record ID
- 189592
- FDA record created
- 2021-10-29
- Recall event number
- 88802