Z-0119-2023: ARTIS pheno, Interventional fluoroscopic x-ray system, Model No. 10849000

Recall record Z-0119-2023 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0119-2023
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA classification date
Oct 21, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Siemens has become aware of three potential software issues with ARTIS pheno systems with software version VE10B. This may lead to a hazardous situation for patients if treatment cannot be continued on the system and treatment needs to be continued on an alternate system.

Source facts

Action
On September 20, 2022, the firm notified affected customers via Urgent Medical Device Correction letters. This notification covered three issues. 1. System Error Management, in which the operator is unable to release X-Ray in "Bypass fluoroscopy" mode which remains permanently inhibited, and may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system. Customers should ensure that patient treatment can be continued in other ways if there is any possible danger for the safety of the patient. 2. Head Holder Symbols, in which the user cannot tell whether a head holder has been selected in the system, may lead to a collision between the C-Arm and the head holder during system movement if the head holder is being used but not selected on the system. Customers should ensure that the configuration of the head holder is correct prior to system movements and double check configuration in the control room in case of uncertainty during the procedure. 3. Stand Movement, in which rarely no stand movement is possible after a routine brake test, leading to a possible delay of procedure. Customers should carry out routine brake tests with sufficient time before starting any procedure. Siemens will correct the software errors and contact customers to arrange a date to perform the corrective action. Please feel free to contact Siemens' service organization for an earlier appointment at 1-800-888-7436.
Initiated
2022-09-20
Root cause
Software design
Distribution
Domestic distribution nationwide. Foreign distribution worldwide.
Recall number
Z-0119-2023
Recall status
Open, Classified
CDRH record ID
196176
FDA classified
2022-10-21
Recall event number
90904

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗