Z-0119-2027: Intellijoint Navigation System: Workstation G2 Configured, REF: 000-0241; Intellijoint HIP Cart System, R

Recall record Z-0119-2027 for Intellijoint Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0119-2027
Company in FDA record
Intellijoint Surgical, Inc.
FDA classification date
Oct 7, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A navigation system used to aid in selection and positioning of orthopedic implant system components has a workstation that may contain a 15-inch MacBook Pro battery, subject to an Apple battery recall, that may overheat and pose a fire safety risk.

Source facts

Action
On 8/28/2026, recall notices were emailed to customers who were asked to do the following: - If affected serial numbers are found, stop use and contact firm's customer service: at customerservice@intellijointsurgical.com and copy regulatory@intellijointsurgical.com - If the observed serial number does not match the affected serial number please record the serial number identified and inform customerservice@intellijointsurgical.com and cc regulatory@intellijointsurgical.com of the correct serial number for that facility.
Initiated
2026-08-28
Root cause
Nonconforming Material/Component
Distribution
Worldwide - US Nationwide distribution in the states of CA, TX, MS, NJ and the countries of Singapore, Spain.
Recall number
Z-0119-2027
Recall status
Open, Classified
CDRH record ID
222550
FDA classified
2026-10-07
Recall event number
99821

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗