Open original FDA record ↗Recall record
- FDA identifier
- Z-0119-2027
- Company in FDA record
- Intellijoint Surgical, Inc.
- FDA classification date
- Oct 7, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A navigation system used to aid in selection and positioning of orthopedic implant system components has a workstation that may contain a 15-inch MacBook Pro battery, subject to an Apple battery recall, that may overheat and pose a fire safety risk.
Source facts
- Action
- On 8/28/2026, recall notices were emailed to customers who were asked to do the following: - If affected serial numbers are found, stop use and contact firm's customer service: at customerservice@intellijointsurgical.com and copy regulatory@intellijointsurgical.com - If the observed serial number does not match the affected serial number please record the serial number identified and inform customerservice@intellijointsurgical.com and cc regulatory@intellijointsurgical.com of the correct serial number for that facility.
- Initiated
- 2026-08-28
- Root cause
- Nonconforming Material/Component
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, TX, MS, NJ and the countries of Singapore, Spain.
- Recall number
- Z-0119-2027
- Recall status
- Open, Classified
- CDRH record ID
- 222550
- FDA classified
- 2026-10-07
- Recall event number
- 99821