Weak true postive/false negative results: weak true positive or false negatives in ruptured fetal membranes may occur due to a "hook effect"
Source facts
Action
AmniSure notified users by Medical Device Correction (Recall) letter dated October 1, 2007 issued via US Mail. The letter provides additional labeling instructions and a checklist for trouble shooting, including a diluation procedure that allows users to determine and solve the problem of the hook effect with this device.
Initiated
2007-10-03
Root cause
Device Design
Terminated
2008-02-06
Distribution
Worldwide, USA, Australia, Austria, Italy, Korea, Bulgaria, Russia, Spain, Kuwait, Switzerland, Luexemburg, Ireland, Denmark, Netherlands, Jordan, Israel, Chile, UK, Singapore, Basil, New Zealand, Turkey, France, Greece, and Equador
Recall number
Z-0120-2008
Recall status
Terminated
CDRH record ID
64550
FDA classified
2008-01-19
Recall event number
44937