Skip to content

Public FDA record

Z-0120-2019: Disposable Smooth Interior Tubing Material: Polyethylene EVA copolymer. This tubing is supplied non-steri

Recall record Z-0120-2019 for A M Systems Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0120-2019
Company in FDA record
A M Systems Inc
FDA record creation date
Oct 16, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Some of the Disposable Smooth Interior Tubing may pull apart when stretched.

Source facts

Action
On 08/10/2018, the firm, A-M Systems, sent an "URGENT: Notification of Product Recall/Correction" letter to customers via email informing them that it was determined that some of the Disposable Smooth Interior Tubing may pull apart when stretched. Customers were advised to provide the firm with the number of items distributed from the affected lots and the number of items still remaining in their inventory from these lots so that the firm could accurately track the status of this correction action. The customers are instructed to please use the enclosed form and self-addressed envelope to return the information to the firm immediately. Once the customers have provided the firm with an accurate count of items distributed and items remaining in the customer's control please destroy or discard any remaining product in your possession from these lot numbers. If you have any questions, please contact Quality Manager at Tel: 360.683.8300 x1302 or Email: rthompson@a-msystems.com.
Initiated
2018-08-10
Root cause
Process control
Terminated
2021-05-13
Distribution
US Distribution to states of: OH, PA, and NC.
Recall number
Z-0120-2019
Recall status
Terminated
CDRH record ID
167392
FDA record created
2018-10-16
Recall event number
81076

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗