Skip to content

Public FDA record

Z-0121-2022: P2131 Automated Field Steam Sterilizer NSN: 6530-01-641-4641

Recall record Z-0121-2022 for Fort Defiance Industries, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0121-2022
Company in FDA record
Fort Defiance Industries, LLC
FDA record creation date
Oct 16, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The rear heater plate and cover assembly that contains the electrical wiring for the heaters does not have sufficient bonding to ground.

Source facts

Action
A Recall notification letter titled, "MEDICAL DEVICE ADVISORY NOTICE" was sent to consignees on 08/19/2021 via a email. The notice provides instructions regarding continued use of the device until FDI provides a protective cover for the rear heater plate and cover assembly. It mentions, "THE OPERATOR SHOULD NOT TOUCH THE REAR HEATER PLATE AND COVER ASSEMBLY OF THE DEVICE WHILE THE DEVICE IS IN USE OR POWERED ON. IF ACCESS TO THE REAR OF THE DEVICE IS REQUIRED, PLEASE TURN THE POWER OF THE UNIT OFF USING THE FRONT PANEL SWITCH OR DISCONNECT INCOMING POWER TO THE DEVICE. DO NOT ATTEMPT TO MODIFY THE DEVICE TO CHANGE THE ELECTRICAL GROUND CONDITION OF THE HEATER PLATE AND COVER ASSEMBLY AS THIS WILL IMPACT DEVICE PERFORMANCE."
Initiated
2021-08-19
Root cause
Device Design
Distribution
Distributed nationwide to PA, MD, TX, NJ, OH, TN, UT, VA and internationally to Canada.
Recall number
Z-0121-2022
Recall status
Open, Classified
CDRH record ID
189120
FDA record created
2021-10-16
Recall event number
88628

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗