Open original FDA record ↗Recall record
- FDA identifier
- Z-0121-2022
- Company in FDA record
- Fort Defiance Industries, LLC
- FDA record creation date
- Oct 16, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The rear heater plate and cover assembly that contains the electrical wiring for the heaters does not have sufficient bonding to ground.
Source facts
- Action
- A Recall notification letter titled, "MEDICAL DEVICE ADVISORY NOTICE" was sent to consignees on 08/19/2021 via a email. The notice provides instructions regarding continued use of the device until FDI provides a protective cover for the rear heater plate and cover assembly. It mentions, "THE OPERATOR SHOULD NOT TOUCH THE REAR HEATER PLATE AND COVER ASSEMBLY OF THE DEVICE WHILE THE DEVICE IS IN USE OR POWERED ON. IF ACCESS TO THE REAR OF THE DEVICE IS REQUIRED, PLEASE TURN THE POWER OF THE UNIT OFF USING THE FRONT PANEL SWITCH OR DISCONNECT INCOMING POWER TO THE DEVICE. DO NOT ATTEMPT TO MODIFY THE DEVICE TO CHANGE THE ELECTRICAL GROUND CONDITION OF THE HEATER PLATE AND COVER ASSEMBLY AS THIS WILL IMPACT DEVICE PERFORMANCE."
- Initiated
- 2021-08-19
- Root cause
- Device Design
- Distribution
- Distributed nationwide to PA, MD, TX, NJ, OH, TN, UT, VA and internationally to Canada.
- Recall number
- Z-0121-2022
- Recall status
- Open, Classified
- CDRH record ID
- 189120
- FDA record created
- 2021-10-16
- Recall event number
- 88628