Skip to content

Public FDA record

Z-0122-06: Siremobil Iso-C Mobile X-Ray System

Recall record Z-0122-06 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0122-06
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA classification date
Nov 2, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Screws may become loose and fall into the OR field during surgery.

Source facts

Action
The recalling firm has their Service Engineers visiting each site to check the screws on the window and secure them with Loctite per Update Instructions SP022/05/S. The Service Engineer is also informing the account of the problem during the visits.
Initiated
2005-08-01
Root cause
Other
Terminated
2006-07-19
Distribution
The products were shipped nationwide to a distributor, a physician, and medical facilities
Recall number
Z-0122-06
Recall status
Terminated
CDRH record ID
41297
FDA classified
2005-11-02
Recall event number
33082

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗