Open original FDA record ↗Recall record
- FDA identifier
- Z-0122-2013
- Company in FDA record
- Orthosensor
- FDA classification date
- Oct 24, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Ortho sensor in Sunrise, FL is recalling the Graphics User Interface Software 1.8 (which was updated to version 1.82, validated on 02/13/12). Recall was initiated due to customer feedback which included the line of site between the knee ba lancer and the Link station was obstructed by OR metal stands or trays or the knee balancer device was placed atop of or near the metal stands or trays, interru
Source facts
- Action
- Orthosensor did not provide a written customer notification. As the recall was reported to the FDA well after all nine of the products were corrected (the recall stemmed from an initial inspection of the firm), then no customer notification was requested. The removal and replacement of the previous version of the Graphic User Interface Software is being handled internally by members of Operations and the Sales force. The items the firm is recalling is software, and the software cannot be returned; the old version (version 1.8) would be erased. The new version 1.82 would be installed and downloaded into the equipment at the site during servicing. This does not involve the customer of which none are consignees. Effectiveness checks associated with the update of the software is being handled internally by the Operations and the Sales force. Further questions please call (813) 645-2855.
- Initiated
- 2012-02-04
- Root cause
- Software design
- Terminated
- 2012-12-12
- Distribution
- Nationwide Distribution including the states of CA, FL, MI and NY.
- Recall number
- Z-0122-2013
- Recall status
- Terminated
- CDRH record ID
- 112490
- FDA classified
- 2012-10-24
- Recall event number
- 63081