Open original FDA record ↗Recall record
- FDA identifier
- Z-0122-2015
- Company in FDA record
- Siemens Healthcare Diagnostics
- FDA classification date
- Oct 23, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Customer complaints were received for positive bias reported on the controls used on the IMMULITE /IMMULITE 1000 on the Cortisol Assay reagent lots IMMULITE /IMMULITE 1000 (LKCO1) Lots 381,382 and 383 . Subsequently, the firm confirmed a positive bias on the IMMULITE/IMMULITE 1000 (LKCO1) Lots 381, 382 and 383.
Source facts
- Action
- US Customers were sent an Urgent Medical Device Recall Letter (UMDR2014-09-19, dated 9/19/14) via FedEx on 9/19/14. Outside US Customers were sent an Urgent Field Safety Notice Letter. The letters inform the consignees of the recall of the IMMULITE/IMMULITE 1000 Cortisol (LKCO1) assay, kit lots 381, 382 and 383. Customers were asked to take the following actions: - Please review this letter with your Medical Director. - Discontinue use of and discard the IMMULITE/IMMULITE 1000 Cortisol kit lots listed in the letter. - Review your inventory of these products to determine your laboratorys replacement needs and to provide information to Siemens for reporting to the authorities. --- Customers are requested to complete and return the Effectiveness Check questionnaire attached to the Urgent Field Safety Notice letter within thirty (30) days.
- Initiated
- 2014-09-19
- Root cause
- Nonconforming Material/Component
- Terminated
- 2015-11-18
- Distribution
- Worldwide Distribution: US (nationwide) and internationally to Canada.
- Recall number
- Z-0122-2015
- Recall status
- Terminated
- CDRH record ID
- 130224
- FDA classified
- 2014-10-23
- Recall event number
- 69321