Skip to content

Public FDA record

Z-0122-2024: Functional MR V1.0

Recall record Z-0122-2024 for OLEA MEDICAL. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0122-2024
Company in FDA record
OLEA MEDICAL
FDA classification date
Oct 18, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When exporting regions of interest in DICOM SEGMENTATION format, when simultaneously exporting multiple volumes of interest, a functional magnetic resonance software bug may cause file contents and associated series descriptions to not match, which could lead to a misinterpretation of clinical data.

Source facts

Action
On 8/10/23, field safety notices were emailed to customers who were told to take the following actions: Advice for the user to detect if the bug occurred: This defect can be detected within the Functional MR V1.0 user interface by displaying the exported volume on an anatomical sequence. To detect the issue, the VOI shall be dragged and dropped from the data browser on an anatomical sequence, which will display a notification, indicating the original VOI s name. Outside of Functional MR V1.0 user interface, the user may check the DICOM tag "Segment Label (0062,0003)" in the viewer used, as this DICOM tag contains the correct VOI s name. Workaround: This problem can be overcome when exporting VOIs individually and has no impact on exporting VOIs merged with an anatomical map. New software will be available in October 2023. Complete and return the reply form via email to qa-ra@olea-medical.com
Initiated
2023-08-10
Root cause
Software design
Distribution
US: MI, MA
Recall number
Z-0122-2024
Recall status
Open, Classified
CDRH record ID
203328
FDA classified
2023-10-18
Recall event number
93049

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗